15
MENU SETUP
1/2
MENU MODE EXPERT
METER TYPE pH
T SENSOR NONE
SET TEMPERATURE 25
CALIBRATION MENU TIMEOUT 05:00
DELAY AT STARTUP 00:05
AUX OUTPUT MODE OFF
YEAR 2013
MONTH 5
DAY 15
HOUR 14
MENU SETUP
2/2
MINUTE 52
SET PASSWORD 0 0 0 0 0 0
SET DEFAULTS NO
MENU MODE: SIMPLE or EXPERT
YEAR: put the current YEAR
MONTH: put the current MONTH
DAY: put the current DAY
HOUR: put the current HOUR
MINUTE: put the current MINUTE
7.3.2. EXPERT MENU MODE
It permits to set the following parameters by use :
MENU MODE: SIMPLE or EXPERT
METER TYPE: pH, Rx, Cl 20ppm,Cl 2ppm, Cl 200ppm, Cl 2000ppm
TSENSOR: NONE or PT100
SET TEMPERATURE: put the current TEMPERATURE is PT100 is not setted
CALIBRATION MENU TIMEOUT: set the TIMEOUT time for CALIBRATION procedure
DELAY STARTUP: set the startup DELAY TIME for the instrument
AUX OUTPUT MODE: set OFF or CLEANING or FLOCCULANT
YEAR: put the current YEAR
MONTH: put the current MONTH
DAY: put the current DAY
HOUR: put the current HOUR
MINUTE: put the current MINUTE
SET PASSWORD: put the password if it is necessary to have it
SET DEFAULTS: set NO or MAINTAIN CALIBRATION or RESTORE ALL