EasyManua.ls Logo

Etatron eSelect B2 - Menu Calibration

Etatron eSelect B2
35 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
24
MENU SETUP
MENU SETPOINT1
MENU SETPOINT2
MENU CALIBRAZTION
MENU ALARM
MENU SETUP
MENU SETPOINT1
MENU SETPOINT2
MENU CALIBRATION
MENU ALARM
MENU CALIBRATION
CALIBRATE pH 7/REF. POINT 8.40
CALIBRATE BUFFER 4/9 8.40
2013/05/15 04:52 WEDNESDAY T=25
METER 3: ALARM: MAX
7.6. MENU CALIBRATION
Starting from the MAIN SCREEN press OK key than press + key to move the selector on MENU
CALIBRATION.
Press OK key to enter.
Press OK key to enter
into MENU
MAIN SCREEN
Chose the parameter by press + or keys.
Press
OK key to confirm the choice and set the
value for the parameter on right column.
Modify the value for the parameter by press + or and
keys. Press
OK
ke
y
to confirm the choice and return
on left column to set an other parameter.
Calibration Menu for pH
mode
620mV
SET1:OFF mA1:10.4
SET1:ON mA2:10.4
7.47 pH
SET1:ON mA1:12.4
SET1:OFF mA2:12.4
11.63 Cl ppm
SET1:OFF mA1:10.4
SET1:ON mA2:10.4

Related product manuals