Setting the bolus dose
You can alter the bolus dose in
increments of 0.1mg/h with the
and
buttons. S
et the desired
bolus dose,
then press .
Set basal rate
You can alter the value in increments
of 0.1 mg/h with the
and
buttons. Set the desired basal rate,
then press
.
Confirm basal rate
Check the daily dose. Press
if the
value displayed is correct.
Press
to return to basal rate
setting.
You can select between min.
0.0 mg/h and max. 15.0 mg/h.
Please refer to the package insert
and observe the maximum drug
dose allowed per hour.
Setting the number of boluses
Set the number of boluses permitted
using the and buttons,
then press
.
It is possible to adjust the
bolus
de
livery amount to between
0.0mg to max. 10.0mg. 0to max.
20boluses possible. Timespan of
6h to 24h possible. Lockout time
of 15min to max. 90min possible.
Setting the time period
Set the desired period of time using
the
and buttons, then press
.
Setting the lockout time
Set the desired lockout time using
the
and buttons, then press
.