88
Eversense XL CGM User Guide
9
Alert Descriptions and Actions (continued)
Alerts
App DisplayDescription
Sensor Check
Appears once when the system’s internal checks
detect instability with the sensor which requires a
return to calibration Initialisation Phase.
Predicted Low Glucose
Appears every 60 minutes when your glucose values
are trending low and will reach your Low Glucose
Alert level within the time you entered in Settings.
Actions
In the Initialisation Phase, you must perform
4 fingerstick calibration tests spaced 2 - 12 hours
apart. Display of glucose readings will resume after
the 2nd successful fingerstick calibration test.
Pay close attention to your glucose values,
symptoms, and trends. Confirm your glucose value
with a blood glucose meter test before making a
treatment decision.