EasyManuals Logo

Excell SI-132 Service Manual

Excell SI-132
60 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #12 background imageLoading...
Page #12 background image
EXCELL PRECISION CO., LTD
02005302 ZSME300000237 11
1-7 Error Messages
Unable to read the EEPROM.
(The EEPROM is not mounted yet, or the circuit near EEPROM is broken.)
Initial zero is higher than the zero range when switching the scale on.
Initial zero is lower than the zero range when switching the scale on.
Initial zero is unstable for 10 seconds. (If appears, please press
ON/ZERO
.)
(OMIL approval models)
Internal value is below Zero.
A/D IC malfunction (cannot read A/D value). The load cell may not be
connected to the indicator correctly.
The weight of the object is over 9 divisions of the maximum capacity.
A/D IC value is over the maximum range.
- - - - -
If the negative weight is over 20 divisions and there is no T or PT, the display
shows “- - - - - -” (OMIL approval models)
1-8 Weight UNIT
(kg) 1 g = 0.001 kg
(g) 1 g = 1 g
(lb) 1 g = 0.002204623 lb
(lb/oz) 1 g = 0.03527396 oz
(oz) 1 g = 0.03527396 oz
(GN) 1 g = 15.432358 GN
(dwt) 1 g = 0.6430149 dwt
(ct) 1 g = 5 ct