EXCELL PRECISION CO., LTD
02005302 ZSME300000237 4
Quick Setup Calibration
This page is to quickly initiate the scale, for the other functions configuration, you can refer the
chapters below.
Instructions:
Step 1 :
¾ Power off the scale, find the switch SWA1 on the bottom of machine.
¾ Switch SWA1 to the ADJ position and then turn the power on. The display will
show
.
Step 2 :
¾ Refer to the chapter 3-1 at page 32 to complete Capacity Setup.
Step 3 :
¾ Refer to the chapter 3-3 at page 41 to complete Weight Calibration.
Step 4 :
¾ Refer to the chapter 3-2 at page 44 to complete Linearity Calibration.
Step 5 :
¾ When done the initiation, switch the switch SWA1 back to the LOCK position.
If the switch SWA1 is switched to the LOCK position during calibration, the machine
will exit the service mode automatically.