EasyManuals Logo

Excell SI-132 Service Manual

Excell SI-132
60 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #47 background imageLoading...
Page #47 background image
EXCELL PRECISION CO., LTD
02005302 ZSME300000237 46
Press
TARE
key to finish
the setting.
The display shows the last
setting. And then key in the
gravity value of manufacture
location.
Manufacture: Taipei
Manufacture: Shanghai
Manufacture: other location
Note1 Note1
3-3-2 Manufacturing Location Gravity Setting
The gravity value should be among the value of Equator and Polar.
Equator gravity G
E
= 9.7803184558 m/sec
2
Polar gravity G
P
= 9.8321772792 m/sec
2
Taipei 9.78914 m/sec
2
Shanghai 9.79423 m/sec
2
Note1:
Use
ON/ZERO
or
UNITS
key
to select the manufacture
location.
.
.
ON/ZERO
key Upward key (0~9 digit entry)
UNITS
key Downward key (0~9 digit entry)
TARE
key Move cursor rightward
NET/GROSS
key Move cursor leftward
Press
TARE
key
Key in “00”
Press
TARE
key
Press
TARE
key
Press
TARE
key
Press
TARE
key