EXCELL PRECISION CO., LTD
02005302 ZSME300000237 21
When the right digit is flashing,
put the platter on the scale and
press
TARE
key while there is
no load on platter.
kg
According to the calibration
weight to put the weight on the
platter and press
TARE
key.
The scale is managing the data
of full capacity. When it
becomes stable, the beeper
makes 3 beep sounds and the
display flickers. Then take off
the weigh to
finish setting
kg
2-2 External Weight Calibration
Conditions for external weight calibration:
The calibration weight value placed on the platter must be over 100e, and the standard
deviation of the weight must be within 10%.
Approval model cannot access
.
.
ON/ZERO
key ⇒ Upward key (0~9 digit entry)
UNITS
key ⇒ Downward key (0~9 digit entry)
TARE
key ⇒ Move cursor rightward
NET/GROSS
key ⇒ Move cursor leftward
Press
TARE
key,
After stable
the display shows the calibration
value (the maximum capacity of 1
st
UNITS setting) and UNITS. The
right digit keeps flashing.
The calibration value could be
changed.
kg
The scale is managing the data of
zero point. When it becomes stable,
the display shows the calibration
weight.