Ex3: 30kg (5kg, 5kg ,10kg, 10kg)
Display Key Press Note
03 CLn TARE Into linearity calibration
L0 TARE First point (zero), remove weigh pan and press the Tare key
L1 1 TARE Put 5kg on and press the Tare key
L2 1 TARE Put 5kg on and press the Tare key
L3 2 TARE Put 10kg on and press the Tare key
L4 2 TARE Put 10kg on and press the Tare key
L5 1 NET/GROSS To finish linearity adjustment (5 points linearity calibration)
03 CLn
Ex4: 30kg (1kg, 2kg ,5kg, 10kg, 2kg, 10kg)
Display Key Press Note
03 CLn TARE Into linearity calibration
L0 TARE First point (zero), remove weigh pan and press the Tare key
L1 1 TARE Put 1kg on and press the Tare key
L2 2 TARE Put 2kg on and press the Tare key
L3 5 TARE Put 5kg on and press the Tare key
L4 A TARE Put 10kg on and press the Tare key
L5 2 TARE Put 2kg on and press the Tare key
L6 A TARE Put 10kg on and press the Tare key
L7 1 NET/GROSS To finish linearity calibration (7 points linearity calibration)
03 CLn
Notes:
In the process of , , press
NET/GROSS
key to abort the linearity calibration.
In the process of , , , , , , or press
NET/GROSS
key to
finish and save the 2, 3, 4, 5, 6, 7, or 8 points calibration.
In the process of , press
TARE
key to finish and save the 9 points calibration.