Remote Control Commands
108 FVA-3100
Specific commands-Quick Reference Command Tree
PROGram:SELected:DURAtion
Description This command changes the duration parameter for the
attenuation program.
Syntax PROG:SEL:DURA<wsp><duration>
Parameters The <duration> parameter is the time that the FVA-3100
remains at each attenuation step. This value has the following
format: “999,59,59” (hours,minutes,seconds).
Example PROG:SEL:DURA 000,10,00
Note This command cannot be issued while a program is running.
PROGram:SELected:STATe
Description This command controls the status of the currently loaded
program.
Syntax PROG:SEL:STAT<wsp><value>
Parameters The <value> parameter can be
“1” or “START” to start the program
“0” or “STOP” to stop the program
Example PROG:SEL:STAT START