Maintenance
Variable Attenuator 63
Cleaning Connectors Equipped with EUI/EUA Adapters
3. Gently wipe the connector and ferrule.
4. With a dry lint-free wiping cloth, gently wipe the same surfaces to
ensure that the connector and ferrule are perfectly dry.
5. Throw out the wiping cloths after one use.
6. After having cleaned the connectors and ensuring that the unit is not
active, verify the surface of the connector with a small, portable
fiber-optic microscope.
IMPORTANT
Isopropyl alcohol takes approximately ten seconds to evaporate.
Since isopropyl alcohol is not absolutely pure, evaporation will
leave microscopic residue. Make sure you dry the surfaces before
evaporation occurs.
WARNING
DO NOT verify the surface of the connector with a fiber-optic
microscope WHILE THE UNIT IS ACTIVE. Doing so WILL result in
permanent eye damage.