EasyManua.ls Logo

Exigo H400 - Guided Calibration

Exigo H400
97 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Page 44 © Boule Medical AB, March 2018. Article no. 1504496
6. Calibration
Calibration
X Guided Calibration
The Guided Calibration is a step-by-step assisted calibration, usable for both Open Tube and MPA
inlets and always requires 5 Boule Calibrator runs.
Note: It is possible to exit the Guided Calibration at any time, and already analysed samples can be
used in Advanced Calibration.
1 Select Mode to be Calibrated Go to Main Menu and login using Authorization Code.
Select Calibration and then choose mode to be calibrated:
Whole Blood
M PA
Select Guided Calibration
2 Background Analysis Operator is guided through a background analysis prior to
start of calibration sequence.
3 Scan in or choose Calibrator
Figure 63: Select or Scan in Calibrator
Dialog is shown prompting to select or scan calibrator to
start analysis 1/5.
4 Calibrator Analysis For OT analysis:
Press Start Plate to aspirate calibrator sample.
(For MPA analysis follow instructions for how to prepare
and run a sample in Chapter 3.)
5 First Results After analysis 1/5, results are presented together with a
textbox stating if analysis is OK or not. If analysis is:
OK: Two but t ons Reject Analysis and Accept Analysis
are shown. If pressing:
Reject Analysis: same proceedure as for Not OK
analysis below occurs
Accept Analysis: New dialog prompting start of
analysis 2/5 appears.
Not OK: A Rerun button is shown, which if pressed
will prompt start of analysis 1/5 again.