EasyManua.ls Logo

Exigo H400 - Page 45

Exigo H400
97 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
© Boule Medical AB, March 2018. Article no. 1504496 Page 45
6. Calibration
Calibration
6 Continuation of Calibrator
Analysis
Figure 64: Consecutive Guided Calibration Analysis
Repeat steps 2 to 4 above until all 5/5 samples have been
analysed.
7 Completion of Guided Calibration
Figure 65: Guided Calibration Results
When pressing button Accept Analysis for sample 5/5 a
calibration result screen is displayed showing results for all
parameters together with a pass/fail status column. Press
button Save to save all new calibration factors (only those
that are indicated pass will be updated).
Note: It is only possible to update and save new Calibration factors for parameters that have
status passed.