Page 70 © Boule Medical AB, March 2018. Article no. 1504496
7. Menu Structure and Advanced Setup
Advanced Parameter Setup
User Login
The instrument allows for two modes of login. Level Login (default), where each level have an unique
password and allows for an increasing set of available settings and functions or User Login, where a
set of personal users are created and each user is assigned a level with its associated set of available
settings and functions. Below is an overview of the Login Levels authorities.
Note: The Admin can change password for all users (Admin does not need to know other users
old passwords). All users can change their own respective passwords (but the old passwords are
required in order to perform this function).
Login Level Description
Basic User
- Analyse, view and search samples
- View sample statistics
- Run Prime
- Manual standby
- View instrument information (About)
- Change own Password
User
Basic User as well as:
- System Setup
- Maintenance
Advanced User
User as well as:
- Advanced Setup
- Calibration
- Deletion of Selected Samples
- Change Login Type (User/Level)
Admin
Advanced User as well as:
- Manage Users
- Change other users Passwords
1
Change Login Type to Old Level Login System
1. Press menu button Login.
2. Set Login Type = Level Login
3. Press button Save to store login type.
2
Change Login Type to New User Login System
1. Press menu button Login.
2. Set Login Type = User Login, and press Accept.
3. Press button Save (an administrator user "Admin" will be
automatically created).
4. Input and then verify the new password for user Admin.
5. Press keypad OK button (resulting in an automatic log out).