WEBMCA Programming
ComTec GmbH 6-4
6.1.11. Preset-Object
PRESET
Start-/Stop conditions
Elements req. Content
START_CONDITION Value (only for date/time)
STOP_CONDITION Value (for date/time, realtime, livetime, counts)
6.1.12. START_CONDITION-Object
START_CONDITION
Attribute(s) req. Content
TYPE X Type of start condition
0 - manual
1 - external via gate signal
2 - via date/time specification
6.1.13. STOP_CONDITION-Object
STOP_CONDITION
Attribute(s) req. Content
TYPE X Type of stop condition
0 - manual
1 - external via gate signal
2 - via date/time specification
3 - Realtime preset
4 - Livetime preset
5 - Counts preset
6.1.14. NETWORK_CONDITION-Object
// TODO
6.1.15. AUTH-Object
// TODO