4.6 Classifications and Ratings
Type of classification Class
Medical Device Directive: IIa (Chair alone: I)
Protection against electric
shock:
Class I
Protection against electric
shock, applied part:
Type B
IP classification: IPX0, No special protection
IP classification of remote Foot
Control:
IPX1, protected against dripping water
Mode of operation: Continuous (Chair alone: non-continuous)
Electrical Rating: 100 V AC, 50/60 Hz, 450 VA (PS150C1/100, PS150C2/100)
110 V AC, 50/60 Hz, 450 VA (PS150C1/110, PS150C2/110)
115 V AC, 50/60 Hz, 450 VA (PS150C1/115, PS150C2/115)
220 V AC, 50/60 Hz, 450 VA (PS150C1/220, PS150C2/220)
230 V AC, 50/60 Hz, 450 VA (PS150C1/230, PS150C2/230)
240 V AC, 50/60 Hz, 450 VA (PS150C1/240, PS150C2/240)
100 V AC, 50/60 Hz, 600 VA (PS2150C1/100, PS2150C2/100)
110 V AC, 50/60 Hz, 600 VA (PS2150C1/110, PS2150C2/110)
115 V AC, 50/60 Hz, 600 VA (PS2150C1/115, PS2150C2/115)
220 V AC, 50/60 Hz, 600 VA (PS2150C1/220, PS2150C2/220)
230 V AC, 50/60 Hz, 600 VA (PS2150C1/230, PS2150C2/230)
240 V AC, 50/60 Hz, 600 VA (PS2150C1/240, PS2150C2/240)
4.7 Information about Electromagnetic Compatibility
The information in this chapter has to be considered when installing and using the F1
product family. Portable and mobile RF communications equipment, for example mobile
phones, can affect the functionality of F1 products.
Guidance and manufacturer’s declaration electromagnetic emissions
The F1 products are intended to be used in the electromagnetic environment specified below. The customer or the user of the F1 product
should assure that it is used in such an environment.
Emissions test
Compliance
Electromagnetic environment guidance
RF emissions
CISPR 11
Group 1
The F1 products use RF energy only for its internal function. Therefore, its
RF emissions are very low and are not likely to cause interference in near-
by electronic equipment.
RF emissions
CISPR 11
Class B
The F1 products are suitable for use in all establishments, including domes-
tic establishments and those directly connected to the public low-voltage
power supply network that supply buildings used for domestic purposes.
Harmonic emissions
IEC 61000-3-2
Class A
Voltage fluctuations/
flicker emissions
IEC 61000-3-3
Compliant