User Guide • 47
Recommended separation distances between portable and mobile RF communications equipment and the F1 products
The Fimet F1 is intended for use in an electromagnetic environment where radiated RF disturbances are controlled. The customer or the
user of the Fimet F1 can help prevent electromagnetic interference by maintaining a minimum distance between portable and mobile RF
communications equipment (transmitters) and the Fimet F1 as recommended below, according to the maximum output power of the com-
output power of
transmitter
W
Separation distance according to frequency of transmitter
m
For transmitters rated at a maximum output power not listed above, the recommended separation distance d in meters (m) can be estimated
using the equation applicable to the frequency of the transmitter, where P is the maximum output power rating of the transmitter in watts (W)
according to the transmitter manufacturer.
NOTE 1 At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from struc-
tures, objects and people.
4.8 Environmental Specifications
Variable Min Max
During storage and transport
Humidity 10% 95%
Temperature -40°C 70°C
Temperatures for display and IDIS -20°C 60°C
Air pressure 50 kPa (0.5 bar) 106 kPa (1.06 bar)
During use
Humidity 30% 75%
Temperature 10°C 35°C
Air pressure 80 kPa (0.8 bar, ca.
2,000 m above sea
level)
102 kPa (1.02 bar, ca.
60 m below sea level)
Using the product in a moving vehicle is prohibited.
4.9 Connections to Networks
Variable Nominal Min Max Min
Volume
Water pressure 150 kPa (1.5 bar) 300 kPa (3 bar) 5 l/min
Air pressure 550 kPa (5.5 bar) 800 kPa (8 bar) 60 l/min
Drainage diameter 25 mm
Suction
50 m3/h 500 mm
H2O
Electricity 230 V AC / 50 Hz
Power 600 W
Current 3 A