FLUXUS G80x A Menu Structure
140 UMFLUXUS_G8V4-6-1EN, 2018-10-10
Current loop
activation of a current output
This display will only be indicated if the current output has been installed in
Special Funct.\SYSTEM settings\Proc. outputs.
selection whether the sign of the measured values is to be considered for the
output
This display will only be indicated if Current Loop is activated.
input of the lowest/highest measured value to be expected for the current out-
put
The values are assigned to the lower/upper limit of the output range.
These displays will only be indicated if Current Loop is activated.
input of the error value delay, i.e. the time interval after which the error value
is transmitted to the output in case no valid measured values are available
This display will only be indicated if Special Funct.\SYSTEM set-
tings\Dialogs/Menus\Error-val. delay = EDIT is selected.
Pulse output
activation of a Pulse Output
This display will only be indicated if a pulse output has been installed in SYS-
TEM settings\Proc. outputs\\.
input of the pulse value (totalizer value at which a pulse will be emitted)
This display will only be indicated if Pulse Output is activated.
input of the pulse width
range: 1...1000 ms
This display will only be indicated if Pulse Output is activated.
Alarm output
Activation of an Alarm Output
This display will only be indicated if an alarm output has been installed in
Special Funct.\SYSTEM settings\Proc. outputs.
Selection of the switching condition (func), the holding behavior (typ) and
the switching function (mode) of the alarm output.
This display will only be indicated if Alarm Output is activated.
selection of the physical quantity to be monitored
This display will only be indicated for R1 if Alarm Output is activated.
input of the upper limit of the physical quantity to be monitored
This display will only be indicated if Alarm Output is activated and MAX is
selected as the switching condition.
input of the lower limit of the physical quantity to be monitored
This display will only be indicated if Alarm Output is activated and MIN is
selected as the switching condition.
INIT-resis-
tant
Meas.Values
>ABSOLUT< sign
Full-Scale Val.
300.00 m3/h
R1=FUNC<typ mode
Function: MAX
R1 Input: ↕
act. Volume flow