FT-112(D) Panel Weighing Indicator, Technical Manual, Rev.1.0.0, May 2019 Page 123 of 170
40159
R/W
2
Decimal point-1
Dec
Description
0
XXXXOO
1
XXXXXO
2
XXXXXX
3
XXXXX.X
4
XXXX.XX
5
XXX.XXX
40161
R/W
2
Increment-1
Dec
Description
0
X 1
1
X 2
2
X 5
40163
R/W
2
MAX-2
40165
R/W
2
Decimal point-2
40167
R/W
2
Increment-2
40169
R/W
2
MAX-3
40171
R/W
2
Decimal point-3
40173
R/W
2
Increment-3
40175
R/W
2
Limit of Indication
Dec
Description
0
Over indication after Max.
1
1 division more than Max.
2
5 division more than Max.
3
9 division more than Max.
4
2% more than Max.
5
5% more than Max.
40177
R/W
2
Tare type
0
Subtractive tare
1
Additive tare
40179
R/W
2
Maximum tare
Refer to par. [526] page 65
40181
R/W
2
Secondary unit
Dec
Description
0
g
(Gram)
1
kg
(Kilogram)
2
t
(Ton)
3
lb
(Libre)
4
No unit
(without unit)
5
N
(Newton)
6
kN
(Kilonewton)
7
kLb
(Kilolibre)
40183
R/W
2
Reserve
40185
R/W
2
Calibration
Commands
Decimal
Definition
0
None
188
Adjust Zero Calibration command
220
Adjust Span Calibration command
(first, load test weight value to 40187)
23205
Apply the coefficients of eCal
(first, load 40189, 40191 and 40193)
40187
R/W
2
Span Calibration Value
40189
R/W
2
Total Load Cell capacity for eCal
eCal
Coefficients
40191
R/W
2
Average mV/V value for eCal
40193
R/W
2
Dead load value for eCal
40195
R
2
Calibration
Process Status
Bit
Definition
B0
Ready for calibration