FT-112(D) Panel Weighing Indicator, Technical Manual, Rev.1.0.0, May 2019 Page 148 of 170
First Unit (power on unit)
Zeroed with power on zero
Actual Weight or
Dynamic Result
Dynamic weight is calculating
Table 15-1– ALL Status table
Calibration Status (always 32 bit integer)
1
st
Dword (input) descriptions when read command is ‘Calibration Status’. Refer to 2
nd
Dword of PLC Output to
FT-112(D) Panel Input
1
st
Dword (input) Description
The Calibration DIP switch is not ‘On’ position.
- Check the calibration DIP switch.
Scale unstable
- Wait until scale become stable
- Check grounding wiring
Calibration load value entry Error
- Test weight is too small. Increase the weight
Calibration Error
- Calibration loading is not enough
- Check test weight loading
- Check load cell connections
Instrument cannot be calibrating
- Load cell signal is very low or too high
Instrument cannot be calibrating
- Check load cell cable
- Re-energize the instrument
ADC Error
- Re-energize the instrument
- If shown again, change the board.
Calibration Timeout
- Restart calibration
Span calibration in process …
Zero calibration in process …
Not ready for calibration
Table 15-2 - Calibration status