Cleaning the camera
23
Note
This section only applies to cameras where removing the lens exposes the infrared
detector.
In some cases the dust cannot be removed by following this procedure: the infrared
detector must be cleaned mechanically. This mechanical cleaning must be carried out
by an authorized service partner.
CAUTION
In Step 2 below, do not use pressurized air from pneumatic air circuits in a workshop, etc., as this air
usually contains oil mist to lubricate pneumatic tools.
Follow this procedure:
1. Remove the lens from the camera.
2. Use pressurized air from a compressed air canister to blow off the dust.
#T810587; r. AA/74078/74713; en-US
97

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the FLIR E96 and is the answer not in the manual?

FLIR E96 Specifications

General IconGeneral
Resolution640 × 480 pixels
Temperature Range-20°C to 1500°C (-4°F to 2732°F)
Field of View24° × 18°
Detector TypeUncooled microbolometer
Digital Camera5 MP
Wi-FiYes
Laser PointerYes
Image Frequency30 Hz
Accuracy±2°C (±3.6°F) or ±2% of reading
Digital Zoom1-8x continuous
Laser Distance MeterYes
Operating Temperature Range-15°C to 50°C (5°F to 122°F)
Storage Temperature Range-40°C to 70°C (-40°F to 158°F)
IP RatingIP54
Thermal Sensitivity<30 mK at 30 °C (86 °F)
Display4" LCD touchscreen
Storage MediaSD card
Battery TypeLi-ion battery
Minimum Focus Distance0.15 m (0.49 ft)
Spectral Range7.5 – 14 μm
Weight1.2 kg (2.7 lbs)