High and Low Alarms
8
The TG series features a high and low temperature alarm with five high/low
set point presets, as explained below.
1. Press and release the trigger to enter the HOLD mode.
2. Long press the MODE button to access the programming menu (Section
9) and view the high alarm set point screen. HI will be shown on the
display.
3. Use the LOG button to select a preset (P1...P5).
4. Use the arrow buttons to set the high alarm set point.
5. Press the MODE button to step to the low alarm limit screen. LO will be
shown on the display. Use the arrow buttons to adjust the low alarm set
point.
6. Pull and release the trigger to exit the programming menu.
Figure 8.1 Setting the high and low alarm set points.
When a temperature measurement exceeds the high or low limit, the HI or LO
icon flashes on the display, in the SCAN or HOLD modes. When measure-
ments are within safe limits (not exceeding an alarm set point), the display
shows OK.
You can set 5 pairs of high/low alarm set points, one pair for each preset (P1...
P5). Step 3, in the procedure above, is where you select a preset.
#NAS100131; r. AC/92852/92852; en-US