Data Logger
10
The TG Series allows you to log readings manually or automatically. There are
two memory banks, one for the manual logger and one for the automatic log-
ger; each bank has 99 locations. After logging readings, you can recall and
delete them per the instructions below. Each logged reading is stored with its
date/time stamp, alarm status, and emissivity setting.
10.1 Manual Datalogger
To store a reading manually, short press the LOG button from either the SCAN
or the HOLD mode. The display will flash twice, confirming that the reading
has been stored. Up to 99 readings can be stored, after which the oldest read-
ings will be overwritten.
To recall or delete logged readings, see Section 10.3 below.
To switch to the automatic data logging memory (Auto Interval) from the man-
ual mode, short press the LOG button. You can then scroll through the read-
ings that were logged in the automatic mode (Auto Interval). See the next
section for Auto Interval details.
10.2 Automatic Data Logger (Auto Interval)
The TG Series allows you to automatically record a selectable number of
readings (up to 99) at a programmable time interval (from 1 minute up to 96
hours and 56 minutes).
Important note: Before you begin setting up the automatic data logger, you
must set the meter’s time and date in the programming menu (Section 9), oth-
erwise the data logger will not operate as expected.
Follow the steps below for configuring and running the automatic data logger.
1. Pull and release the trigger to enter the HOLD mode.
2. Long press the MODE button to access the programming menu.
3. Short press the MODE button seven times to step to the Auto Interval
menu.
4. Continue with the steps in Table 10.1. below.