Specifications
12
12.2 IR Thermometer Specifications
Range TG54–2: -22 to 1562℉ (–30 to 850℃)
TG56–2: -22 to 2372℉ (–30 to 1300℃)
Out-of-range indication OL (over range) and —OL (under range)
Resolution 0.1° maximum
Accuracy
Accuracy specifications apply for
ambient temperature conditions
(21 to -25℃ [69.8 to 75.2℉]) and
with an emissivity setting of 0.95.
-22 to 14℉ (-30 to -10℃): ±7.2℉ (±4.0℃)
14 to 32℉ (-10 to 0℃): ±4.5℉ (±2.5℃)
32° to 212℉ (0° to 100℃): ±1.8℉ (±1.0℃)
> 212℉ (100℃): ±1%
Spectral Response
8 to 14 microns
Temperature Coefficient
±0.1°
Distance to Spot (D:S) ratio 20:1 for TG54–2; 30:1 for TG56–2
Response time TG54–2: 250 ms
TG56–2: ≤ 250 ms
Emissivity setting 0.01 to 1.00 with 5 presets
12.3 Laser Specifications
Laser type Single point Class 2 laser
Wavelength 650 nm (±20 nm)
Power <1mW
12.4 Thermocouple Specifications (TG56–2)
Type Type K
Meter display range for thermo-
couple measurements
-22° to 1202℉ (-30 to 650℃)
Supplied thermocouple range -22 to 500℉ (-30 to 260℃)
Do not attempt to measure temperature > 500℉
(260℃) with the supplied thermocouple.
Resolution 0.1°
Accuracy Probe: ±3.0℉ (±1.5℃)
Meter: ±1.8℉ (±1℃) or 1% (whichever is greater)
#NAS100131; r. AC/92852/92852; en-US
31