EasyManua.ls Logo

FLIR TG54-2 - IR Thermometer Specifications; Laser Specifications; Thermocouple Specifications (TG56-2)

FLIR TG54-2
42 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Specifications
12
12.2 IR Thermometer Specifications
Range TG54–2: -22 to 1562 (–30 to 850)
TG56–2: -22 to 2372 (–30 to 1300)
Out-of-range indication OL (over range) and —OL (under range)
Resolution 0.1° maximum
Accuracy
Accuracy specifications apply for
ambient temperature conditions
(21 to -25 [69.8 to 75.2]) and
with an emissivity setting of 0.95.
-22 to 14 (-30 to -10): ±7.2 (±4.0)
14 to 32 (-10 to 0): ±4.5 (±2.5)
32° to 212 (0° to 100): ±1.8 (±1.0)
> 212 (100): ±1%
Spectral Response
8 to 14 microns
Temperature Coefficient
±0.1°
Distance to Spot (D:S) ratio 20:1 for TG54–2; 30:1 for TG56–2
Response time TG54–2: 250 ms
TG56–2: 250 ms
Emissivity setting 0.01 to 1.00 with 5 presets
12.3 Laser Specifications
Laser type Single point Class 2 laser
Wavelength 650 nm (±20 nm)
Power <1mW
12.4 Thermocouple Specifications (TG56–2)
Type Type K
Meter display range for thermo-
couple measurements
-22° to 1202 (-30 to 650)
Supplied thermocouple range -22 to 500 (-30 to 260)
Do not attempt to measure temperature > 500
(260) with the supplied thermocouple.
Resolution 0.1°
Accuracy Probe: ±3.0 (±1.5)
Meter: ±1.8 (±1) or 1% (whichever is greater)
#NAS100131; r. AC/92852/92852; en-US
31