Loading...
9 - 12 92B/96B/99B/105B Users Manual
Multimeter mode (V dc, excluding probe factor):
Table 5
Table 6
Table 7
Susceptibility: no visible disturbance
Frequency range:
10 kHz to 25 MHz
E = < 0.1V/m E = 1 V/m E = 3 V/m
Stand alone 100 mV to 300V 100 mV to 300V 100 mV to 300V
With PM8918 100 mV to 300V 100 mV to 300V 100 mV to 300V
Susceptibility: no visible disturbance
Frequency range:
25 MHz to 1 GHz
E = < 0.1V/m E = 1 V/m E = 3 V/m
Stand alone 100 mV to 300V 100 mV to 300V 100 mV to 300V
With PM8918 100 mV to 300V 300 mV to 300V 3V to 300V
Susceptibility: disturbance less than 10 % of full scale
Frequency range:
25 MHz to 1 GHz
E = < 0.1V/m E = 1 V/m E = 3 V/m
Stand alone - - -
With PM8918 - 100 mV 100 mV to 300 mV