Appendixes 9 - 11
Table 3
Table 4
Susceptibility: disturbance less than 10% of full scale
Frequency range:
10 kHz to 25 MHz
E = < 0.1V/m E = 1 V/m E = 3 V/m
Stand alone - - -
With PM8918 - - 1 mV/div to 2 mV/div
Susceptibility: disturbance less than 10% of full scale
Frequency range:
25 MHz to 1 GHz
E = < 0.1V/m E = 1 V/m E = 3 V/m
Stand alone - 1 mV/div 1 mV/div to 2 mV/div
With PM8918 - 50 mV/div to 200 V/div 1 V/div to 100 V/div
For conditions not specified in tables 1 to 4, a susceptibility effect of more than 10 % is possible