Insulation Tester
Measurements
23
View Results Stored in Memory
Note
Parameters not appropriate for a test are shown as INVALID.
The Tester can store 99 sets of test data, including:
Tags
Ramp on or off
Insulation Resistance
Timer reading at termination of test (Timer)
Test Voltage Selected (TV)
Actual Test voltage (V)
Capacitance I
Polarization Index (PI)
Dielectric absorption ratio (DAR)
Test current (I)
Reason for ending the test
Limit – off or timer setting (1 to 99 minutes) (T. Limit)
To view stored test data, see Figure 11:
1. With the Tester turned on, push to call the Function menu.
2. Push or to select 3. Show Results.
3. Push to select the menu item.
Note
When a voltage is present at the terminals, that voltage is always
shown on the top-center of the display, regardless of whether that
voltage is sourced by the Tester or is from the circuit under test.