1550C/1555
Users Manual
30
Electromagnetic
Compatibility
EN 61326-1, EN 61326-2-2
Certifications P ® )
Safety Compliance EN 61010-1, EN 61557 Parts 1 and 2
CAT III 1000V, CAT IV 600V
Pollution Degree 2
Test Voltages Number of Tests
250 V 4100
500 V 3600
1 kV 3200
2.5 kV 2500
5 kV 1000
Typical Battery Charge
Capability
Note
At temperature extremes, the
battery needs to be charged
more frequently.
10 kV 500
Electrical Specifications
The Tester’s accuracy is specified for 1 year after calibration at operating
temperatures of 0 °C to 35 °C. For operating temperatures outside the range (-
20 °C to 0 °C and 35 °C to 50 °C), add ±0.25 % per °C, except on the 20 %
bands add ±1 % per °C.
Insulation
Test Voltage (DC) Insulation Resistance Range Accuracy (±reading)
250 V < 200 kΩ
200 kΩ to 5 GΩ
5 GΩ to 50 GΩ
> 50 GΩ
unspecified
5 %
20 %
unspecified
500 V < 200 kΩ
200 kΩ to 10 GΩ
10 GΩ to 100 GΩ
> 100 GΩ
unspecified
5 %
20 %
unspecified