Fluke 190 Series II
Users Manual
114
Automatic Scope Measurements
The accuracy of all readings is within ± (% of reading +
number of counts) from 18 °C to 28 °C. Add 0.1x (specific
accuracy) for each °C below 18 °C or above 28 °C. For
voltage measurements with 10:1 probe, add probe
accuracy, see section ‘10:1 Probe’ on page 127. At least
1.5 waveform period must be visible on the screen.
General
Inputs.............................................................A, B, C and D
DC Common Mode Rejection (CMRR) .................. >100 dB
AC Common Mode Rejection at 50, 60, or 400 Hz ...... >60 dB
DC Voltage (VDC)
Maximum Voltage
with 10:1 probe ..................................................... 1000 V
direct (1:1)............................................................... 300 V
Maximum Resolution
with 10:1 probe ........................................................ 1 mV
direct (1:1)............................................................. 100 µV
Full Scale Reading............................................. 999 counts
Accuracy at 5 s to 10 μs/div
2 mV/div......................................... ±(1.5 % + 10 counts)
5 mV/div to 100 V/div ........................ ±(1.5 % + 5 counts)
Normal Mode AC Rejection at 50 or 60 Hz ............. >60 dB
AC Voltage (VAC)
Maximum Voltage
with 10:1 probe...................................................1000 V
direct (1:1) ............................................................300 V
Maximum Resolution
with 10:1 probe......................................................1 mV
direct (1:1) ..........................................................100 µV
Full Scale Reading ..........................................999 counts
Accuracy
DC coupled:
DC to 60 Hz................................ ±(1.5 % +10 counts)
AC coupled, low frequencies:
50 Hz direct (1:1)....................... ±(1.5 % + 10 counts)
60 Hz direct (1:1)....................... ±(1.9 % + 10 counts)
With the 10:1 probe the low frequency roll off point will
be lowered to 2 Hz, which improves the AC accuracy
for low frequencies. When possible use DC coupling
for maximum accuracy.
AC or DC coupled, high frequencies:
60 Hz to 20 kHz......................... ±(2.5 % + 15 counts)
20 kHz to 1 MHz........................... ±(5 % + 20 counts)
1 MHz to 25 MHz........................ ±(10 % + 20 counts)
For higher frequencies the instrument’s frequency roll
off starts affecting accuracy.
Normal Mode DC Rejection.............................>50 dB

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Fluke 190 Series II and is the answer not in the manual?

Fluke 190 Series II Specifications

General IconGeneral
BrandFluke
Model190 Series II
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals