Remote Operations
Remote Commands 3
3-9
CALibration:SECure:PASSword <spd>
Description Sends a password to the Product to put the Product into adjustment mode.
Note
The calibration switch on the rear panel of the Product must be set to
Enable before you send the password.
Parameters <spd> = (the password)
Example
CAL:SEC:PASS “ADJUST” Sends the string ADJUST to the Product to
allow calibration adjustments. Note the word
ADJUST must be in quotes for this
command.
CALibration:SECure:EXIT
Description Exits the calibration or adjustment mode.
Example
CAL:SEC:EXIT Exits calibration adjustments.
Note
All changes will be lost the next time the Product is power cycled, unless
CAL:SAVE? is done before the exit command.
CHAin:FITTed?
Description Gets the number of units in the chain.
Example
CHA:FITT? Get back a number from 1 to n.
CHAin:IDN? <nrf>
Desription Gets the identification string of an instrument in the chain.
Parameter <nrf> = 1 to 10 Index of the instrument in the chain.
Example
CHA:IDN? 2 Gets back the identification string of the second
instrument in the chain.
Note
The command
:CHA:IDN? 1 is equivalent of *IDN?.
Returns <manufacturer>, <model>, <serial number>, <firmware>
INPut:TYPE(?) <cpd>
Description Sets or gets the input type selection.
Parameter <cpd> = CURRent Sets the input type to current.
VOLTage Sets the input type to voltage.
Example
INP:TYPE CURR Sets the input type to current.
INP:TYPE? Gets back CURR or VOLT depending on which
type is set in the Product.
OUTPut[:STATe](?) <bool>
Description Sets or gets the state of the output signal.
Parameters <bool> = ON or 1 Enables the output signal (OPR).
OFF or 0 Disables the output signal (STBY).
Example
OUTP ON Turn the Product output on.