EasyManua.ls Logo

Fluke 5790B - Page 118

Fluke 5790B
208 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
5790B
Operators Manual
6-28
CAL_CLST?
Lists the symbolic names and values for a group of calibration constants.
Parameters
1. Which constants to list. This parameter may be a single constant (for
example, DI_7V) or a group of constants (for example, AC_7V), or a
group of groups (for example, AC, or ALL). Refer to Appendix B for a
listing of calibration constants and group names, or use the command "
CAL_RPT? CONSTS" to print a list of all calibration constants.
2. ACTIVE, STORED, OLD, or DEFAULT.
Response
(String) A list of the selected calibration constant names and values. The form
is:
"<EOL>
<total><EOL>
<name>,<value><EOL>
<name>,<value><EOL>
etc.
CAL_CONST?
Returns the value of a particular calibration constant from the active set.
Parameter
The symbolic name of the calibration constant (for example, DI_7V). Refer to
Appendix B for a listing of calibration constant symbolic names, or use the
command " CAL_RPT? CONSTS" to print a list of all calibration constants.
Response
(Float) A number signifying the value of the constant.
CAL_COUNT?
Sequential command. Returns the number of times calibration constants have
been saved at the end of a calibration procedure (except for DC Zeros)
Response
integer
Parameter
None
Example
CAL_COUNT?
Returns 34 if calibration constants have been saved 34 times.

Table of Contents

Other manuals for Fluke 5790B

Related product manuals