EasyManua.ls Logo

Fluke 5790B - Page 120

Fluke 5790B
208 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
5790B
Operators Manual
6-30
CAL_DAYS?
Returns the number of days elapsed since the most recent verification or
calibration.
Parameters
No argument (Returns the number of days since the most recent
verification, as set by CAL_DATE)
MAIN (Entire main input calibration)
WBND (Wideband calibration)
SERVICE (Service calibration)
DC (DC calibration)
AC (AC calibration)
ZC (DC zeros calibration
WDC (Wideband 1 kHz calibration)
WAC (Wideband Flatness)
ALL (All of the above)
A group name from Appendix C, for example,
AC_2_2VResponseResponse
(Integer) The number of elapsed days.
Response
(Integer) The number of elapsed days.
CAL_DC
Begins a dc calibration procedure for one range or all ranges. As in CAL_AC, this
specifies which step the next CAL_NEXT command initiates.
Parameters
1. INPUT1 or INPUT2 (The input to be used; signifies dc calibration), or
WBND for Option 5790B/3 or Option 5790B/5 Wideband gains calibration
2. (Optional) A number that defines the range amplitude (same as for the
RANGE command). If this parameter is included, do only the specified
range; if not included, do all ranges.
Example
An example sequence of commands for calibrating a range (in this case, the
220 mV range) is as follows:
CAL_TEMP 23.5
CAL_DC INPUT2, 200 mV
CAL_NEXT 200 mV
CAL_NEXT -200 mV
An error occurs if either the entered value is out of range or the measured
input is out of range.

Table of Contents

Other manuals for Fluke 5790B

Related product manuals