Remote Commands
Alphabetical List of 5790B Device-Dependent Commands 6
6-57
SPLSTR
Sets the serial/USB/Ethernet remote mode Serial Poll response string.
Parameters
The string (up to 40 characters) to print on receipt of a ^P (Control/P)
character.
SH
Load or create a shunt. If a shunt of that serial number is already stored, it is
loaded and the value is not checked.
If no shunt of that number is stored, a generic shunt of the type is created with
the given serial number.
If the instrument is not in the current shunt measurement window, the window will
be enabled on the successful execution of this command.
Parameter
1. Shunt Serial Number
2. Shunt Value
The following is a list of valid shunt values:
CUSTOM, SH1MA, SH10MA, SH20MA, SH50MA, SH100MA, SH200MA,
SH500MA, SH1A, SH2A, SH5A, SH10A, SH20A, SH50A, or SH100A
SH?
Returns the data for a current shunt.
SH? Returns the data for the selected shunt.
SH? <sernum> Returns the data for a stored shunt with the given serial number.
Parameter
Serial number of a stored shunt. (Optional - If omitted, shows the data for
the currently selected shunt). A serial number of a stored shunt.
Example
SH?
val=100A; sernum=000000000 (MODIFIED); caldate=0; R(nom)=8.000e-03 ohms;
DCerr=0 ppm
8 AC/DC err pts:
+0 ppm @ 1.000000e+01 Hz
+0 ppm @ 5.500000e+01 Hz
+0 ppm @ 4.000000e+02 Hz
+0 ppm @ 1.000000e+03 Hz
+0 ppm @ 1.000000e+04 Hz
+0 ppm @ 3.000000e+04 Hz
+0 ppm @ 7.000000e+04 Hz
+0 ppm @ 1.000000e+05 Hz
5 loading err pts:
+0 ppm @ 0.000000e+00 Hz
+0 ppm @ 1.000000e+03 Hz
+0 ppm @ 1.000000e+04 Hz
+0 ppm @ 3.000000e+04 Hz
+0 ppm @ 1.000000e+05 Hz