Remote Commands
Alphabetical List of 5790B Device-Dependent Commands 6
6-61
SH_DELE
Clears the shunt data for a particular serial number.
The generic shunt for each class cannot be cleared (nor can serial number 0).
Parameter
1. (Integer) Serial Number of shunt to delete
Example
SH_DELE 000134445
Deletes the data for shunt serial number 000134445. Use the SH_LISTALL?
command for a complete list of shunt values and their serial numbers.
SH_LDGEN
Loads the generic shunt data for a value of shunt.
Parameter
CUSTOM, SH1MA, SH10MA, SH20MA, SH50MA, SH100MA, SH200MA,
SH500MA, SH1A, SH2A, SH5A, SH10A, SH20A, SH50A, or SH100A
Example
SH_LDGEN SH20MA
Loads the generic shunt data for the 20MA current shunt
SH_LIST?
Lists the serial numbers available for a value of shunt.
Parameter
CUSTOM, SH1MA, SH10MA, SH20MA, SH50MA, SH100MA, SH200MA,
SH500MA, SH1A, SH2A, SH5A, SH10A, SH20A, SH50A, or SH100A
Response (comma separated list)
1. (Integer) Number of shunts for the selected value of shunt
2. (Integer) Serial number
.
.
.
Example
SH_LIST? SH20MA
2,4562,1345 (lists 2 20 MA shunts with serial numbers 4562 and 1345)