Introduction and Specifications
Secondary Electrical Specifications 1
1-17
Voltage
Range
Frequency Range
24 Hour AC
Stability ± 1 °C
Slow Filter Peak-
Peak ± μV
Temperature Coefficient
[1]
Input
Resistance
[2]
10 °C to 40 °C
0 °C to 10 °C
40 °C to 50 °C
ppm / °C
700 mV
10 Hz - 20 Hz 8 1.5 3.0
20 Hz - 40 Hz 6 1.5 3.0
40 Hz - 20 kHz 6 1.5 3.0 10 MΩ
20 kHz - 50 kHz 6 2.0 3.0
50 kHz - 100 kHz 12 5.0 8.0
100 kHz - 300 kHz 18 10.0 10.0
300 kHz - 500 kHz 36 20.0 20.0
500 kHz - 1 MHz 96 50.0 50.0
2.2 V
10 Hz - 20 Hz 8 1.5 3.0
20 Hz - 40 Hz 5 1.5 3.0
40 Hz - 20 kHz 5 1.5 3.0
20 kHz - 50 kHz 5 2.0 3.0 10 MΩ
50 kHz - 100 kHz 10 5.0 8.0
100 kHz - 300 kHz 18 10.0 10.0
300 kHz - 500 kHz 30 20.0 20.0
500 kHz - 1 MHz 90 50.0 50.0
7 V
10 Hz - 20 Hz 8 1.5 3.0
20 Hz - 40 Hz 5 1.5 3.0
40 Hz - 20 kHz 5 1.5 3.0
20 kHz - 50 kHz 5 2.0 3.0 50 kΩ
50 kHz - 100 kHz 10 5.0 8.0
100 kHz - 300 kHz 18 15.0 15.0
300 kHz - 500 kHz 30 30.0 30.0
500 kHz - 1 MHz 90 65.0 65.0
22 V
10 Hz - 20 Hz 8 1.5 3.0
20 Hz - 40 Hz 5 1.5 3.0
40 Hz - 20 kHz 5 1.5 3.0
20 kHz - 50 kHz 5 2.0 3.0 50 kΩ
50 kHz - 100 kHz 10 5.0 8.0
100 kHz - 300 kHz 18 15.0 15.0
300 kHz - 500 kHz 30 30.0 30.0
500 kHz - 1 MHz 90 65.0 65.0
70 V
[3]
10 Hz - 20 Hz 8 1.5 3.0
20 Hz - 40 Hz 5 1.5 3.0
40 Hz - 20 kHz 5 1.5 3.0
20 kHz - 50 kHz 5 2.0 3.0 50 kΩ
50 kHz - 100 kHz 18 5.0 8.0
100 kHz - 300 kHz 36 15.0 15.0
300 kHz - 500 kHz 48 40.0 40.0
500 kHz - 1 MHz 120 75.0 75.0