Front Panel Operation
Instructions for Voltage Measurements 4
4-23
One example of a Transfer mode application is determining the error of an ac
source referenced to an external dc standard such as a Fluke 5730A Calibrator.
In transfer mode, an external source can be used to establish a reference, and
the uncertainty of the measurement is governed by Product ac-dc difference
specifications in addition to the uncertainty of the dc source.
Before putting these procedures into practice, see Techniques for Reducing
Transfer Error. Observing the hints provided there will help ensure accurate
transfers. See Figures 4-1 and 4-2 for test lead connection guidelines. (In these
figures, a multifunction calibrator is used as the ac and dc reference.)
+/- 22.0 ppm(90 Days) 9 days since verification
2/06/15 10:52am
Back
Clear
Reference
Average
Reference
View
Reference
Select
Formula
(Vin-Avg)/Avg = +5.3 ppm
Input 2
V
kHz
1.0000
2.000003
Auto Range 2.2 V
hvi038.eps