Fluke 8010A - Probe Accessories Specifications; H Touch-Hold Probe 600-2; AC Current; Probe 600-2

Fluke 8010A
73 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
ACCESSORIES
PROBE
ACCESSORIES
Table 600-1.
Probe
Accessories Specifications
80T-H
TOUCH-HOLD PROBE
Voltage Ratings
TIP
TO
COMMON . 1000V
dc or
peak ac
maximum
COMMON
TO GROUND
. 60V dc
or
42V peak ac maximum
Input
Capacitance
.
150
pF
maximum
Lead
Resistance
. 0.50 maximum
80T-150C
AND 80T-150F TEMPERATURE
PROBES
Range
(«C/°F)
.
50®
C to -FI
50°
C/-58° F to
302°
F
Accuracy
.
±1°C
(1.8°
F)
from
0°Cto
100°C,
decreasing
linearly to
±3°
C
(5.4°
F)
at-50°C
and
+150°C
Sensitivity
1mVdc/°Cor°F
Resolution .
0.1 °C
or
0.1°
F
recommended
maximum
Voltage
Standoff :
.
350V
dc or peak ac
Power
.
Internal
disposable
battery;
1.000
hours of
continuous use
801-400
AC
CURRENT
PROBE
Range , 1A to 400A
Accuracy
.
±{3%
+ 0.4A)
48
Hz to
440
Hz
±(4%
-F 0.4A) 440 Hz
to 1000 Hz
Output
.
1 milliamp
per ampere
of
input current
(1
mA/A)
Working
Voltage
. 660V ac rms
maximum
Maximum
Conductor
Size
. 1 ea. 30 mm
(1.18") or
2
ea. 25 mm
(0.98")
Shunt (load)
Resistance
. Less than
12
ohms
at
multimeter input to
maintairi specified
accuracy
Typical Bandwidth
.
-10%
at 40 Hz and
5 kHz (1A,
400 Hz reference,
excludes
multimeter
response)
Usable
Current Range
. 0.1A
to
600A,
5 seconds
maximum
above 400A
Safety
. Protection
Class ii
as defined
in lEC
348
and ANSI C39.5
801-600
AC
CURRENT
PROBE
Range
.
1
to
600A
ac
Accuracy
.
±3%
Frequency
Response .
30
Hz
to
1 kHz,
10 kHz typical
Division
Ratio . 1000:1
Working
Voltage
. 750V rms
maximum
Insulation
Dielectric Withstand Voltage
.
5
kV
Maximum
Conductor
Size
. 2-inch diameter
80J-10
CURRENT SHUNT
Shunt
. 10 amps
at 100 mV
Accuracy
(18X
to
28‘»C)
DC
TO 10
KHZ
. ±0.25%
10
KHZ-100
KHZ .
Rising
to
1
dB at 100 kHz typical
Temperature
Coefficient
.
0.005%/°
C
inductance
. 8.3 nH
in
series w/O.OIQ
shunt
Overload
.
Up to
one minute
at 20A with
a
1/4
duty cycle for
recovery
after currents between
10A and 20A
Connects
to
.
3/4 inch
center
banana jacks
Connectors
. 5-way
binding posts
(red and
black)
600-6