EasyManua.ls Logo

Fluke 8010A - 601-11. Line Frequency Noise Rejection; 601-13. Theory of Operation (Rechargeable Battery Option); 601-15. Maintenance (Rechargeable Battery Option); 601-20. AC Line Voltage Selection

Fluke 8010A
73 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
OPTIONS
MAINTENANCE
601-11. Line
Frequency
Noise
Rejection
601-12.
Noise
rejection can
be
selected for 50 Hz
or
60
Hz
line
frequencies.
The
line
frequency, as
set at
the
factory,
is
indicated
on a
decal
attached
to the bottom
of
your
Multimeter.
Refer to
the
Line
Frequency
Selection
procedure,
given
later
in
this
section,
to change the
noise
rejection
frequency,
601-13.
THEORY
OF
OPERATION
601-14.
The
theory
of
operation
of
the Multimeter is
illustrated
by
the
Main
PCB
Assembly
schematic in
Section
7.
The
battery
power
supply
is
shown
above
the
standard
power
supply
on
the
schematic. The
Multimeter
can
be
used
with
line
voltage
from
90V to 264V,
50
Hz or
60
Hz
(To
select
the
proper
line
voltage
configuration,
refer
to
the
AC
Line
Voltage
procedure,
given later in
this
subsection.)
Fuse F3
provides
protection for
the
power
supply.
Line
power
input
is
rectified,
filtered, and
regulated.
The
output
of
the
power
supply
acts
as a
current
source
for
the
battery.
The
battery
determines
the
voltage
level
into
the
power
converter. Do
not
operate
your
Multimeter
with
the
battery
removed.
The
power
converter
uses
the
flyback
transformer
technique
to
develop
several
output
voltages
so
that +1
1.5,
-11.5,
+2.9
and
-6.5V
(with
respect
to
power
supply
common)
are
available.
601-15.
MAINTENANCE
WARNING
THESE
SERVICING
INSTRUCTIONS
ARE
FOR
USE
BY
QUALIFIED
PERSONNEL
ONLY.
TO
AVOID
ELECTRICAL
SHOCK, DO
NOT
PERFORM
ANY
SERVICING
OTHER
THAN
THAT
CONTAINED
IN
THE
OPERATING
INSTRUCTIONS
UNLESS YOU
ARE
QUALIFIED
TO
DO SO,
601-16.
Battery
Replacement
601-17.
Use
the
following
procedure
for removing
and
replacing
batteries;
1.
Complete
the
Calibration
Access
procedure
given
in
Section
4 of
this manual.
2.
Turn
the
instrument
upside
down.
3.
Unplug
the
red and
black battery
wires from
the
pcb pins.
4.
Apply
pressure
to
the
front and
rear
sides of
the
7.
Reconnect
the
battery
case
to
the
Main PCB
Assembly.
8.
Connect
the
battery
wires
to their
appropriate
pins
on
the main
pcb,
with- red
to +
and
black
to
-.
9.
Reinstall
the
instrument
in
its
case.
601-18. Fuse
Replacement
601-19.
Use
the
following
procedure
to
replace the main
power
fuse, F3:
1.
Complete
the
Calibration
Access
procedure
given
in
Section
4 of
this
manual.
2. F3
is
located
immediately
in
front
of the
power
receptacle.
Replace
F3 with
a’ 1/32A,
250V,
type
MDL
fuse.
601-20. AC Line
Voltage
Selection
601-21. Use
the
following
procedure to
change
the
operating
acTine
voltage
on the
8010A-01
or 8012A-01:
NOTE
If
the
change
in
operating
voltage
also
changes
the
line frequency
refer
to the
Line
Frequency
Selection
procedure,
given
later in
this
section.
1.
Complete
the
Calibration
Access
procedure
given
in
Section 4
of this
manual.
2.
Refer
to
Figure
601-1
for
the
location of the ac
line
switch
S2,
on
the
Main
PCB
Assembly.
3.
Set S2
to
the 1
lOV
or
220V
position depending
upon
the
intended
line
voltage.
4.
Reinstall
the
case
and
relabel the
ac line
voltage
designation
on the
decal at
the bottom
of the
instrument.
601-22.
Line
Frequency
Selection
601-23.
Use
the
following
procedure to
change
the
line
frequency
noise
rejection
on
the
8010A-01
or
8012A-01:
1.
Complete
the
Calibration
Access
procedure
given
in
Section
4 of
this
manual.
battery
case
to
disconnect the
case from
the
Main
PCB
Assembly.
5.
Remove
the
blotting
paper.
6.
Replace
the
batteries
and
blotting
paper
with
a
new
Fluke
battery
assembly.
601-2
2.
Refer to
Figure
601-1 for the
location of
the line
frequency
selection
switch
S3,
on
the Main
PCB
Assembly.
3.
Set
S3
to the 50
Hz or 60
Hz
position
depending
upon the
intended
line
frequency.

Table of Contents

Other manuals for Fluke 8010A

Related product manuals