EasyManua.ls Logo

Fluke 8010A - 601-24. Power Supply Check

Fluke 8010A
73 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
OPTIONS
POWER
SUPPLY
CHECK
Figure
601-1.
Line Switch (S2) Location
601-24.
POWER
SUPPLY CHECK
601-25.
Use
the
following procedure to
verify the
proper
power
supply voltages of the
8010A-01
or
8012A-01:
1 .
Remove the
ac line
power
cord from the UUT.
If
necessary, charge the battery of
your
Multimeter
before removing the cord.
2.
On the
UUT,
set the
POWER
switch to
ON
position.
4.
Verify
a display reading
of
“00.0”
+0.1V.
5.
With the DMM, verify
the
power supply
voltages
for each test
point
(TP) listed in Table
601-1. The voltages
given under
the heading
Voltage
Levels
are with respect to
ground,
unless
otherwise
indicated.
(Ground
is
provided at TP6.)
601-26. LIST
OF
REPLACEABLE PARTS
601-27. Table
601-2
lists the replaceable
parts
for
the 8010A-01
and 8012A-01 Multimeters.
Figures
601-1 and
601-2 show the component locations
for
8010A-01
and
8012A-01
Multimeters.
3.
On
the
UUT,
select the
DC V
function, 200V
range.
601-3

Table of Contents