ACCESSORIES
PROBE
ACCESSORIES
Table
600-1.
Probe Accessories
Specifications (cont)
80K-6 HIGH
VOLTAGE PROBE
Voltage
Range
0
to 6 kV, dc
or
peak
ac
Input
impedance 75
MO
(nominal)
Division
Ratio
1000:1
Accuracy
DC
TO 500 HZ
±1%
500
HZ TO 1
KHZ
±2%
,
ABOVE
1 KHZ
Output
reading falls. Typically,
30%
at 10
kHz.
80K-40
HIGH
VOLTAGE PROBE
Voltage
Range
1
kV
to 40
kV
dc
or
peak ac, 28
kV rms
ac
Input
Resistance
1000 MO
Division
Ratio
1000:1
Accuracy
Dc (overall) 20
kV to
30
kV
±2%
(calibrated at 25
kV)
UPPER
LIMIT
Changes linearly from
2%
at
30
kV
to 4% at 40
kV
LOWER
LIMIT
Changes
linearly from
2%
at 20 kV to 4% at 1 kV
Accuracy
Ac
(Overall)
±5%
at
60
Hz
83RF HIGH
FREQUENCY
PROBE
AC-to-DC
Ratio
Ratio Accuracy
(at
1
MHz
and loaded
with 10
MQ)
ABOVE IV
BELOW IV
Frequency
Response
(Relative
to
1 MHz)
Extended
Frequency
Response
Response
Voltage Range
Maximum Input
Voltage
Input
Capacitance
Temperature
Range
OPERATING
STORAGE
Humidity
Output
Connector
Accessory
±1 dB
±1.5 dB
100 kHz to 100 MHz
±
1 dB
Useful for
relative readings from 20 kHz to 250 MHz
Responds
to the peak value
of an
input
and is calibrated to read
rms
value
of
a
sine wave
0.25 to 30V rms
30V rms. 200V dc
Approximately
3
pF
+10°Cto+35°C
-40°
C to
+75°
C
<90% R.H.
Fits standard 0.75-inch
dual
banana
connectors
BNC
to
Probe Adapter. P/N 574756
85RF
HIGH FREQUENCY PROBE
AC-to-DC Ratio
1:1
Ratio Accuracy
(At
1 MHz and loaded
with 10 mO
ABOVE 0.5V
±0.5 dB
BELOW
0.5V
±1.0
dB
Frequency
Response
(Relative to 1 MHz) 100
kHz
to 100 MHz ±.5 dB
*100
MHz to 200 MHz
±1.0 dB
*200
MHz
to
500
MHz
±3.0
dB
‘Referred to high and low inputs at probe tip.
Extended
Frequency Response Useful for relative readings from 20 kHz to
700
MHz
Response
Responds to the peak value of an
input
and is
calibrated to
read rms
value of a sine wave
Voltage
Range
0.25 to
30V
rms
Maximum
Input Voltage 30V rms,
200V
dc
600-7