8508A
Users Manual
Clear Calibration Memory
CLRMEM
ALL
phs
HFTRIM
Clear Calibration Memory
CLRMEM
Clears the calibration correction memories.
Caution
This command can obliterate the results of an expensive original
calibration or re calibration!
The extent of clear is defined by programming the following options:
ALL
Clears all
HFTRIM
Clears AC HF frequency response correction only.
Execution Errors
If special calibration is not enabled.
adj289f.eps
4-52