Range Zero and Gain Points
1. On the 8508A, press the Plus key followed by the Tru to select the required
range.
2. On the Calibrator select zero and
OPR to operate the output.
Alternatively if using a standard resistor, connect the standard resistor for a 4 wire
resistance zero.
3. On the 8508A press
CAL to enter the CALIBRATION menu.
4. Press
SAMPLE to initiate the adjustment operation.
Adjustment is complete when the busy legend goes out and the calibrated
measurement is displayed.
5. On the Calibrator select the range gain resistance value.
Alternatively if using a standard resistor, re-connect the standard resistor to measure
it's resistance.
6. If the Set feature is being used:
On the 8508A, select Set on the
CALIBRATION Menu.
Use the numeric keys with the
SET VALUE menu to enter the true output value of the
calibrator at the range gain point, then press Enter.
7. Press
SAMPLE to initiate the adjustment operation.
Adjustment is complete when the busy legend goes out and the calibrated
measurement is displayed.
8. Press the Plus key followed by Truto revert to the range menu.
Tru
LoI Mode Adjustment
Follow the procedure for Normal Mode Adjustment except select the Plus key
followed by Tru instead of selecting the key. Press the CONFIG key and select LoI.
Range Zero and Gain Adjustment Procedure
After the initial setup and connecting up, use the following procedure to adjust range zero
and range gain for the ranges and points detailed in Table 6-5 for the Tru LoI mode.
Just one range can be adjusted if required, but for a full adjustment start with the 2
range and work up to the 20 k range.
On each range, the 8508A automatically recognizes the appropriate value as range zero or
range gain from the applied resistance value.
Note
It is unlikely that the resistance standard will be at exactly the nominal value
required. The Set feature allows a user to enter the true output value of the
calibration standard where it differs from nominal range gain point. If the
resistance value is exactly the nominal value required, omit step 6 for the
Range Gain Point.