Adjustment and Verification
Performance Verification
Table 6-9. AC Voltage Verification Points and Sequence
8508A Range Voltage Frequency 8508A Filter Tolerance
1
200 mV 100 mV 20 Hz 10 Hz ± 0.0195 mV
100 mV 55 Hz 40 Hz ± 0.0175 mV
100 mV 1 kHz 40 Hz ± 0.0149 mV
100 mV 3 kHz 40 Hz ± 0.0175 mV
100 mV 10 kHz 40 Hz ± 0.0175 mV
100 mV 30 kHz 40 Hz ± 0.0445 mV
100 mV 60 kHz 40 Hz ± 0.0995 mV
100 mV 100 kHz 40 Hz ± 0.0995 mV
2 V 1 V 20 Hz 10 Hz ± 0.000144 V
1 V 55 Hz 40 Hz ± 0.000119 V
1 V 1 kHz 40 Hz ± 0.000099 V
1 V 3 kHz 40 Hz ± 0.000119 V
1 V 10 kHz 40 Hz ± 0.000119 V
1 V 30 kHz 40 Hz ± 0.000295 V
1 V 60 kHz 40 Hz ± 0.000795 V
1 V 100 kHz 40 Hz ± 0.000795 V
1 V 500 kHz 40 Hz ± 0.034000 V
1 V 1 MHz 40 Hz ± 0.034000 V
20V 1 V 1 kHz 40 Hz ± 0.00032 V
10 V 20 Hz 10 Hz ± 0.00144 V
10 V 55 Hz 40 Hz ± 0.00119 V
10 V 1 kHz 40 Hz ± 0.00099 V
10 V 3 kHz 40 Hz ± 0.00119 V
10 V 10kHz 40 Hz ± 0.00119 V
10 V 30 kHz 40 Hz ± 0.00295 V
10 V 60 kHz 40 Hz ± 0.00795 V
10 V 100 kHz 40 Hz ± 0.00795 V
10 V 500 kHz 40 Hz ± 0.34000 V
10 V 1 MHz 40 Hz ± 0.34000 V
19 V 1 kHz 40 Hz ± 0.00167 V