6
Adjustment and Verification
Performance Verification
Table 6-12. AC Current Verification Points and Sequence
8508A Range Current Frequency Tolerance
1
200 µA 100 µA
305 Hz
± 0.0830 µA
100 µA
1 kHz
± 0.0830 µA
100 µA
3 kHz
± 0.0830 µA
100 µA
5 kHz
± 0.0830 µA
100 µA
10 kHz
± 0.0830 µA
2 mA 1 mA 305 Hz
± 0.000580 mA
1 mA 1 kHz
± 0.000580 mA
1 mA 3 kHz
± 0.000580 mA
1 mA 5 kHz
± 0.000580 mA
1 mA 10 kHz
± 0.000580 mA
20 mA 10 mA 305 Hz
± 0.00580 mA
10 mA 1 kHz
± 0.00580 mA
10 mA 3 kHz
± 0.00580 mA
10 mA 5 kHz
± 0.00580 mA
10 mA 10 kHz
± 0.00580 mA
200 mA 100 mA 305 Hz
± 0.0545 mA
100 mA 1 kHz
± 0.0545 mA
100 mA 3 kHz
± 0.0545 mA
100 mA 5 kHz
± 0.0545 mA
100 mA 10 kHz
± 0.0545 mA
2 A 1 A 305 Hz
± 0.000945 A
1 A 1 kHz
± 0.000945 A
1 A 3 kHz
± 0.001055 A
1 A 5 kHz
± 0.001055 A
1 A 10 kHz
± 0.001055 A
20 A 10 A 305 Hz
± 0.01140 A
10 A 1 kHz
± 0.01140 A
10 A 3 kHz
± 0.02740 A
10 A 5 kHz
± 0.02740 A
10 A 10 kHz
± 0.02740 A
1. Based on 365 day Tcal ±1°C 99 %CL Absolute specifications. See comments in Applicable Verification Tolerance section above