EasyManua.ls Logo

Fluke PM6690 - Page 85

Fluke PM6690
200 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Troubleshooting 6-41
must be replaced if the normal control voltage range cannot
make the oscillator output 10.000000 MHz.
As a last resort to exclude external causes of malfunction,
desolder the oven oscillator from the main circuit board.
Place it upside down and connect +12 V and ground accord
-
ing to Figu re 6-68. A cold oven oscillator draw s approx i
-
mately 0.30 - 0.35 A. During heating the current consumption
varies. After 10 minutes it should stabilize on less than 0.1 A.
The output V
ref
should be approximately +5 V and the 10
MHz sinewave output signal should have an amplitude of
more than 2.5 V
pp
measured with a 1 MW, 10x probe. The
control input has an internal bias to keep the output frequency
in the middle of the range. Adjust the control voltage between
0 V and +5 V and check the output frequency range with a fre
-
quency counter. The minimum trimming range should be
±5 Hz. 10.000000 MHz must be reached somewhere between
0Vand+5V.
If the oven oscillator circuitry is repaired, a new calibration
must be perform ed. See Chapter 7. A new factory calibration
by means of the utility program should also be performed.
External Reference Input
See Figure 6-67 and Figure 6-70.
The input signal is amplified in U31. The output signal from
the amplifier should be a square wave with logic levels, repro-
ducing the timing characteristics of the input signal. Check
the signal at U 32:11. U32 generates a short pulse (approxi-
mately 40 ns) f or each input cycle, check at U32:9. These
pulses generate a broad spectrum o f harmonics, and the
following high-Q 10 MHz crystal filter allows only a 10 MHz
sinewave to pass. Measure at X19. Note that the trimmer
C442 is used for maximizing the amplitude at X19. Check that
the amplitude is not less than 1 V
pp
. If external reference is not
selected, the gate U33 stops the 10 MHz signal. The control
signal on U33:1 is then low.
100 MHz Multiplier
See Figure 6-67 and Figure 6-69.
100 MHz is used in the measuring logic, mainly as a reference
clock, but also for other purposes. A PLL is used for multi
-
plying the 10 MHz reference to 100 MHz. On power -up the
processor sets up the PLL IC (U9) via the SPI bus. An output
signal, PLL LOCK, tells the processor if the loop is locked
(high level). A VCO, consisting of an inverter (U47) and an
LC circuit in the feedback loop, is controlled by the PLL IC.
The DC voltage from U9:2 is filtered and controls a capaci
-
tance diode. The VCO frequency changes with the capaci
-
tance. The loop can handle the switching of 10 MHz refer
-
ence, from internal to external and vice versa. There is n o
need for a new setup. If external reference is selected and no
such signal is connected to the instrument, the PLL will be un
-
FPGA
100 MHz
PLL
EXT REF
IN
INT REF
OUT
STD
OSC
OVEN
OSC
U4
U4-
U7
10 MHz
U11
ON/OFF
100 MHz
PLL
LOCK
(to mP)
ON/OFF
mP/SPI
mP
m
P/SPI
U41, Q53, Q54
U31, U32,
U33, Q55
U9, U47,
U48
Figure 6-67 Timebase reference system.
3
2
15
4
+12 V
Vref
Vcontrol
GND
10 MH
z
OUT
Figure 6-68 Oven oscillator pinning (seen from bottom side).

Table of Contents

Other manuals for Fluke PM6690

Related product manuals