Calibration Manual
12
R
INS
Insulation Resistance Method I (No Voltage)
To v e r i f y :
1. Set all equipment and the DUT to standby mode.
2. Connect shielded test leads from DUT blue connector to the 532xA LO Port and DUT green
connector to the 532xA HI Port.
3. Connect shielded test leads between DUT red connector and DUT blue connector.
4. Disconnect all other test leads. See Figure 5.
Figure 5. R
INS
Insulation Resistance Method I (No Voltage) Connections
5. Turn the rotary switch on the DUT to R
INS
.
6. Push
button to select R
INS
I mode.
7. Set the 532xA calibrator to output the first resistance test point and set the voltage output
of the DUT to the first voltage setting (see Ta b le 5 ).
ANALYZER
Fluke 5320A
Fluke SMFT-1000
LO
METER
,
L
,
L
+/-