Multifunction PV Analyzer
Performance Tests
29
8. Record the indication on the DUT in the test record.
9. Set the 552xA calibrator to output the next voltage test point (see Ta b l e 1 1 ).
10. Repeat steps 7 through 9 for all test points.
11. Set the 552xA calibrator to standby mode.
AC/DC Current (SMFT only)
To veri f y :
1. Set all equipment and the DUT to standby mode.
2. Connect low-thermal EMF test leads (see Figure 12):
a. Connect DUT green connector to NORMAL red port on the 552xA and DUT yellow
connector to the 552xA black port under NORMAL red port.
b. Disconnect all other test leads.
3. Turn the rotary switch on the DUT to P
AC/DC
.
4. Push
until Current shows in highlight on the left-side menu.
Figure 12. AC/DC Current Connections (SMFT Only)
5. Set the 552xA calibrator to output the first current test point (see Ta b l e 1 2 ).
Note
The 552xA output is voltage, not current, for this part of the calibration.