Calibration Manual
8
11. Momentarily push
on the DUT.
Wait for the DUT to finish both the + and - measurements. Record the + and - indications on
the DUT in the test record. Also record the nominal value displayed on the 532xA as it may
be slightly different from the intended nominal. If applicable, recalculate the limits based on
the nominal value and the published specifications.
Note
Ignore the red X or green
. They do not indicate a pass or fail for this calibration.
12. Set the 532xA calibrator to output the next resistance test point (see Ta b l e 2 ).
13. Repeat steps 10 through 12 for all test points.
14. Set the 532xA calibrator to standby mode.
V
OC
(Open-Circuit Voltage)
To v e r i f y :
1. Set all equipment and the DUT to standby mode.
2. Connect low-thermal EMF test leads from DUT red connector to 552xA NORMAL port and
DUT blue connector to the 552xA black port below NORMAL port. Disconnect all other test
leads (see Figure 3).
Figure 3. V
OC
(Open-Circuit Voltage) Connections
3. Turn the rotary switch on the DUT to V
OC
/I
SC
.
POWER
0
1
23
456
7
89
ENTER
M
k
m
VHz
FIELD
EDIT
/
+
F
OPR EARTH SCOPEEXGRD
MENU
PREV
SHIFT
RESET
CE
SETUP
REF
NEW
TC
MEAS
F
n
p
p
W
dBm sec
C
A
MULT
x
DIV
MODES
MORE
STBY
5522A
CALIBRATOR
A
N
A
L
YZER
Fluke 5522A
Fluke SMFT-1000

Other manuals for Fluke SMFT-1000

Related product manuals