SMFT-1000
Calibration Manual
16
R
INS
Insulation Resistance Method I (With Voltage)
To v e r i f y :
1. Set all equipment and the DUT to standby mode.
2. Connect shielded test leads from DUT red connector to the 532xA LO Port and DUT green
connector to the 532xA HI Port.
3. Connect shielded test leads between DUT red connector to the + voltage output of the
2290-5 and DUT blue connector to the - voltage output of the 2290-5.
Figure 6. R
INS
Insulation Resistance Method I (With Voltage) Connections
4. Turn the rotary switch on the DUT to R
INS
.
5. Push
to select R
INS
I mode.
6. Set the 532xA calibrator to output the first resistance test point and set the voltage output
of the DUT to the first voltage setting (see Ta b le 6 ).
ANALYZER
Fluke 5320A
Fluke SMFT-1000
Power Supply
LO
METER
,
L
,
L
+/-