SMFT-1000
Calibration Manual
22
R
INS
Insulation Resistance Method II (With Current)
To v e r i f y :
1. Set all equipment and the DUT to standby mode.
2. Connect shielded test leads from DUT blue connector to the 532xA LO Port and DUT green
connector to the 532xA HI Port.
3. Connect high-current shielded test leads from DUT red connector to the Output Terminal +
of the current supply (Keithley 2260B-30-36) and from DUT blue connector to the Output
Terminal - of the current supply.
4. Disconnect all other test leads. See Figure 8.
Figure 8. R
INS
Insulation Resistance Method II (With Current) Connections
5. Turn the rotary switch on the DUT to R
INS
.
6. Push
to select R
INS
II mode.
7. Set the 532xA calibrator to output the first resistance test point and set the voltage output
of the DUT to the first voltage setting (see Ta b le 8 ).
A
N
A
LYZE R
Fluke 5320A
Fluke SMFT-1000
LO
METER
,
L
,
L
+/-
2260B